When designing a new catheter / sheath / delivery system tubing, the plurality of parameters can leave you lost in the process. Here's an easy method to jump over this entry barrier into your first iteration — cheat! Or in other words, reverse engineering.

When designing a new catheter / sheath / delivery system tubing, the plurality of parameters can leave you lost in the process. For example, after considering all other interfaces, you will need to select wall thickness, materials, number of layers, lubricants, braid / coil reinforcement, surface treatments and more — where to start?

I'm suggesting an easy method to jump over this entry barrier into your first iteration — cheat! Or in other words, reverse engineering.

First — Set Geometric Anchors

Set geometric anchors to your design. Usually ID and wall thickness are a good place to start.

Second — Get Reference Components

Get hold of any components similar to your design. You can start with raw materials in on-line stores (Nordson MEDICAL, Chamfr and Spectrum Plastics Group are great examples). Other options are sourcing used / expired products in the market (Medical Materials, Inc.) or contact the MedTech Secrets community members for scraps and samples.

These may be expensive, so make sure to extract all information possible from each article; start with non-destructive evaluation, measuring all geometric features such as lengths, diameters and material transitions. Move to functional testing like tensile or bend strength, sheathing forces or even assembling a purpose-built prototype. Now you are ready to carefully tear apart the component under microscope and take note of all investigative results like bond methods, post processes or lumen path tracking.

Medical device opened into two halves showing internal mechanical components

Careful teardown of a reference device under measurement

Third — Document Everything

I cannot stress enough the value of documenting your findings. I suggest to experiment with different documentation tools (I use Microsoft OneNote) to find the simplest way and avoid losing valuable hard-earned information.

In the following example, a braided sheath was measured, dissected and interrogated. Eventually I cut a piece of tube and flattened it to take the picture below: blue thermoplastic outer jacket, SS braid pattern, and a milky white thermoplastic inner liner.

Close-up of a blue thermoplastic tube with a cutout revealing internal metallic braid reinforcement

Flattened tube sample showing outer jacket, SS braid, and inner liner

As the design and performance suited my application, we wanted to figure out the material of this milky polymer. We suspected the liner is Arkema Pebax but needed the exact durometer, and direct measurement wasn't possible as it is bonded to the jacket (and we don't own a durometer). But temperature can be used to differentiate, as each Pebax grade has a slightly different melting point seen in the matrix below.

Shore hardness and density data table for various Pebax grades

Pebax grade differentiation by shore hardness and melting point

Now we faced the problem of isolating the white liner to measure the melting temperature. Fortunately a tip I received from Calvin Vang solved the problem in an elegant manner — a soldering iron! The iron can be gradually and accurately heated to each grade's melting point. Once the liner melts to the touch of the tip, the material is finally revealed.

To Conclude

When making a first tubing iteration of a new design:

  • Set basic design anchors, like ID or OD
  • Cheat! Research references of similar design
  • Relentlessly document findings
#Reverse Engineering#Catheter Design#Tubing#Pebax#Materials